What is your opinion..

What is your opinion measuring CVP thru Power PICC SOLO catheters with optimal tip positions and a patent lumens. The waveform is often lower in amplitude but is the pressure accurate?  

lynncrni
 No, the integral valve built

 No, the integral valve built into the catheter hub would prohibit the pressure measurement. Lynn

Lynn Hadaway, M.Ed., RN,  CRNI

Lynn Hadaway Associates, Inc.

PO Box 10

Milner, GA 30257

Website http://www.hadawayassociates.com

Office Phone 770-358-7861

dcole
Bard makes an adaptor that

Bard makes an adaptor that bypasses the valve in order to get accurate waveforms.   Contact you rep for more details.

 

Darilyn

nurseirie
I know there is a

 

I know there is a adapter out there.  Its not available i Sweden. My Bard rep. claims the cvp is accurate. Im doubtful. Its ok, maybe, if you follow the cvp trend.

Nurse specialist ICU/ANE/PICC.