I need to verify if line is now with tip in the Right Innominate, is this still considered Central? During my line audits I found that line is out by 2 cm, the floor is infusing Calcium Chloride PRN.
Debra Rivie VAT (newbie)
Long Beach Memorial Medical Center
Long Beach, CA
Definitely NOT central and this is considered a malpositioned CVAD. And infusing a horrible vesicant. A serious complication just waiting to happen. See INS standard on CVAD Malpositions. Lynn
Lynn Hadaway, M.Ed., RN, CRNI
Lynn Hadaway Associates, Inc.
PO Box 10
Milner, GA 30257
Website http://www.hadawayassociates.com
Office Phone 770-358-7861