Hello Everyone,
I was wondering if anyone has any evidence based data to support the amount of time the end cap for a VAD should be scrubbed with alcohol or CHG when accessing the lumen (central or peripheral IV's). I'm seeing many policies and procedures that require a 15 second scrub. This is the equivalent of singing "Happy Birthday" twice.Â
I have tried to introduce this expectation at our facility with great resistance and reasonably so if there isn't any evidence to suggest that this amount of time would indeed make a difference.
I appreciate anything you could provide (I already have the Kaler article).
Thanks Very Much!
Janet Mulligan
Nursing Director IV Therapy Service Mass General Hospital Boston
The Kaler article is what everyone is using. To my knowledge there are no other similar studies that have looked at this. Unfortunately the Kaler study did not assess a shorter time period. I have heard Dr. Maki at various presentations state that 10-15 seconds should be what is used. Sorry, this is what we have to work with right now. I would also hasten to emphasis that cleaning is a critical component, so please do not back down on your emphasis of this important step in eliminating CRBSI.
Lynn Hadaway, M.Ed., RN, BC, CRNI
www.hadawayassociates.com
Lynn Hadaway, M.Ed., RN, CRNI
Lynn Hadaway Associates, Inc.
PO Box 10
Milner, GA 30257
Website http://www.hadawayassociates.com
Office Phone 770-358-7861
Did you check out the discussion in other forum threads - Kaler being one of the search options?
Mari Cordes, BS RNIII VA-BC
Vascular Access Department
University of Vermont Medical Center
New research is on the way! I have been in progress for the last 2-3 yrs with a program of research on a new disinfection device designed to disinfect injection ports, needleless connectors, catheter hubs, and stopcocks. This research will be presented in the poster session at SHEA this month. Further research is in progress.
While access site disinfection may seem relatively simple, there are actually multiple variables that play into the complete eradication of organisms on a surface and within catheter hubs or stopcocks. These include the disinfecting agent, the concentration of the agent, the volume of the agent, the contact time (dependent on the first three), the method of application, and the innoculum of microorganisms (even more difficult to eradicate if in biofilms). All have to work together!
Kaler's study demonstrated success with a 15 sec application (measured by flush through), however this was with only an innoculum of 2 log of bacteria (>100cfu, <1000 cfu) on the connector surfaces, some of which have low bacterial tranfer rates (FDA requires minimum 3 log measure for ingress testing for needleless connectors, now requiring 4 log reduction for antimicrobial treated connectors). It may take longer for eradication when higher innoculums are present, which based on the literature is usually the case. Based on my findings of extensive biofilms on connectors retrieved from patient connectors in an ICU (poster presented at APIC 2008) we are dealing with fairly high innoculums (within biofilms).
Kaler did not indicate drying time. Alcohol prep pads do not leave much residual and dries quickly but CHG prep pads are very saturated (a very good thing as they are designed for skin antisepsis). I have found a visible volume of CHG left on the Smart Site Plus and the UltraSite after use of the CHG prep pads which takes up to 2 minutes to dry. This is probably of no consequence in adults? but may be in neonates (I will be evaluating this as well)
Also remember that the site of colonization associated with risk for bloodstream infection in catheter hubs and stopcocks is in the internal surfaces of the hub and not the outside threads (although good to clean this area as well, especially if blood is present. It is virtually impossible to disinfect the inside of a catheter hub or stopcock with a prep pad. How would you do that? You would probably contaminate it in the process especially since you automatically contaminate both sides of a prep pad on removal from the package touching both sides. Catheter hub disinfection protocols are (and stopcock disinfection should be) a requirement of the Joint Commission (thank goodness as this is a missing link to getting to zero for CRBSI) as well as a SHEA/IDSA recommendation.
Happy to share the poster after presentation at SHEA if there is interest.
Marcia Ryder
One other thing.......be careful when using the verbage "scrub the hub" in care campaigns. What hubs are you talking about? Be sure that you differentiate between injection ports, needless connectors (access sites) and catheter/stopcock "hubs". Needless connectors are injection ports, hubs are hubs of the catheter and stopcock hubs. They are different and both now require disinfection and surveillance measures.
Marcia Ryder
Marcia,
Would you post your poster presentation from SHEA?
Thank you,
Dee Dee
Yes, Dee Dee. I will do that. The poster is large with a lot of content so that if I attach the 8 1/2 x 11 version you won't be able to read it when you print it out. So I am having it split in 3 sections so you can print and put together.
Thanks for asking.
Marcia
Marcia,
Did you post this? I looked for it and couldn't find it!
Thanks, Cheryl
Cheryl Kelley RN BSN, VA-BC
Dear Cheryl and Dee Dee,
I have attached the poster as requested. It has been reformated into pages so that it is easy to print out and read. Happy to address any questions you may have.
Next level research is ongoing.
Marcia Ryder
Marcia, I have tried several different methods but can not open your attachment. Any suggestions? Lynn
Lynn Hadaway, M.Ed., RN, BC, CRNI
www.hadawayassociates.com
Lynn Hadaway, M.Ed., RN, CRNI
Lynn Hadaway Associates, Inc.
PO Box 10
Milner, GA 30257
Website http://www.hadawayassociates.com
Office Phone 770-358-7861
Hello Lynn,
Thanks for letting me know.
It is a pdf file. I have resaved the file, let me know if this works.
Thanks marcia
I got it this time but still had to drag to my desktop and open from within Adobe Acrobat. There are several extension on the file name - .pdf.txt.webloc - Strange. Thanks for posting this.
Lynn Hadaway, M.Ed., RN, BC, CRNI
www.hadawayassociates.com
Lynn Hadaway, M.Ed., RN, CRNI
Lynn Hadaway Associates, Inc.
PO Box 10
Milner, GA 30257
Website http://www.hadawayassociates.com
Office Phone 770-358-7861
Thank you Lynn. Sorry for the hassle.....I sometimes have this problem with aol
marcia
Kathy Kokotis
Bard Access Systems
Marcia thanks for sharing the research. We are going to see a huge change in this market this year and next. With new swab solutions and new caps
kathy
Kathy Kokotis
Bard Access Systems