Forum topic

4 posts / 0 new
Last post
Carol Gosselin
peripheral blood Tx via infusion pump

I have been asked if there is any evidence based data as to whether one can infuse blood safely via a peripheral vein using an infusion pump. tradionally at our hospital we have not, the concern being that the pump would not alarm soon enough if the blood infiltrated. What are others doing and are there any articles or studies pro or con? Thanks as always for your Help.

Carol Gosselin Miriam Hospital Providence RI.

lynncrni
Go to your hospital's blood
Go to your hospital's blood bank in the lab and asked to borrow the Technical Manual from the American Association of Blood Banks. This large textbook has a chapter on blood administration techniques. Also, you should check with your pump manufacturer as they have several documents to share. Lynn

Lynn Hadaway, M.Ed., RN, NPD-BC, CRNI

Lynn Hadaway Associates, Inc.

PO Box 10

Milner, GA 30257

Website http://www.hadawayassociates.com

Office Phone 770-358-7861

rivka livni
Speaking only about the

Speaking only about the adult patient, Some IV pumps should not be used for blood infusion, because they cause damage to the RBC as the pump "rolls" over the tubing. It depends on the pump, you should check the manufactor recomendations.

If your pump and tubing are OK to use, it should be OK then. Blood infiltration, though not good for the patient, should be absorbed over time without significant residual effect if caought in time. Our institution has frequent checks during transfusion, so you cathch it before it gets out of hand.  It is very similar to a large bruise you get, when your vein "spills over" after an injury, as far as I know.

Gwen Irwin
Are your pumps FDA approved

Are your pumps FDA approved for blood transfusions?  And the tubing that fits those pumps?  We transfuse blood and blood products (yes, platelets, too) on pumps all the time.

I haven't seen peristaltic pumps in forever (those would be the ones that roll over the tubing).  YOu should ask for manuals for the pumps from the manufacturer for the documentation about FDA approved for blood transfusion.

Most pumps on the market today have a very low psi setting that indicates something is wrong with the amount of pressure that the pump is using to push fluid/blood.  Usually, an occlusion alarm is what is indicated on the pump's screen.  This can be caused by crimped tubing, occluded cathteter, or infiltrated PIV site.  If troubleshooting and tubing isn't bent, catheter is open, then look at the IV site to assess infiltration.  (This should prevent extravasation of blood to a minimal amount.  I would also your manufacturer about the psi exerted by your pumps.)

 Gwen

 

Log in or register to post comments