Q-Syte: Need the good and bad

Has anyone converted to the BD Q-Syte Luer Access Split Septum cap system and want to share the good and bad?  Do you use on ALL vascular access devices?

Thanks for any input.

lynncrni
The good - a split septum

The good - a split septum system with reports of lower rates of infection according to 2 posters presented at AVA last year. 

 The bad - definitely requires the positive pressure flushing technique to prevent blood reflux and clotting. 

My experience comes from the patient's family side when my sister was in an Atlanta hospital that uses these. In fact, this hospital was one who did one of those AVA posters. They use this on all catheters. She did have one PIV that was clotted and had to be restarted. Don't have any experience from a management standpoint about conversion. 

Lynn Hadaway, M.Ed., RN, BC, CRNI

www.hadawayassociates.com

Lynn Hadaway, M.Ed., RN, NPD-BC, CRNI

Lynn Hadaway Associates, Inc.

PO Box 10

Milner, GA 30257

Website http://www.hadawayassociates.com

Office Phone 770-358-7861

Marilyn Hanchett
Suggest you look at work by

Suggest you look at work by or speak to Marcia Ryder RN PhD. Her biofilm analysis of the split septum design reflects some serious infection risk issues. She did a great presentation at this year's APIC, including lab photos of split septum before & after use. Definitely worth your time to check out.

Marilyn Hanchett RN

momdogz
Also try to get a hold of

Also try to get a hold of the ECRI Health Devices article on Needleless Connectors.  Q-Syte was listed as one of the less desirable connectors.  One of the reasons included that it would allow access with a needle.

We looked at the Q-Syte as part of the Nexiva system, and found that if the connector had any pressure on the side from just the right angle, the top would tilt off leaving a wide gap...leaking, and contamination.  It looked like a patient with an administration set attached to their IV site could fairly easily cause this to happen. 

 

Mari Cordes, BS RN 